Alarms
indicator in the alarm adjustment dialog. Alarm guard limits are set in the
Care Unit
Settings
and they are password protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
Setting arrhythmia alarms
You can set the arrhythmia alarms in the
Alarms Setup
menu, or in the
ECG
menu.
1.
Select
Alarm Setup
from the monitor’s main menu.
2.
Select the
Arrhythmia
tab.
3.
Select
Lethal Alarms
.
You can now select the
Alarm Priority
,
Create Snapshot
, and
Print on Alarm
options per arrhythmia.
4.
If Full Arrhythmia license is enabled, you can also select options for the
Atrial
Alarms
and
Ventricular Alarms
.
●
Ventricular Alarms
: You can select the
Alarm Priority
,
Create Snapshot
, and
Print on Alarm
options.
●
Atrial Alarms
: You can select the
Alarm Priority
,
Create Snapshot
, and
on Alarm
options. In addition, you can set the detection criteria for
SV Tachy
:
SVT Length
,
HR for SVT /min
, and
Pause Interval
.
Selecting parameter alarm priority levels
Escalating an alarm priority increases the priority of the alarm condition or increases
the sense of urgency of an alarm signal. The alarm priority is based on clinical
considerations.
The allowed priorities for different alarm groups are defined in the
Care Unit Settings
and they are password protected.
For more information, see the CARESCAPE Modular Monitors Supplemental
Information Manual.
1.
Select
Alarm Setup
from the monitor’s main menu.
2.
Select the
Alarm Priorities
tab.
3.
Select the alarm group:
ECG
,
Invasive Pressures
, or
Other Parameters
.
4.
Select the alarm and its priority from the list.
Selectable alarms are:
2062971-001
CARESCAPE Modular Monitors
141
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...