Monitoring basics
●
Alarm signals are working and can be seen and heard in your care environment.
●
Required parameter calibrations are completed.
Supply mains interruption
If the supply mains to the equipment is interrupted for less than 15 minutes (and the
B650 is used without the optional battery), the monitor keeps the trend data and
the latest user-made settings. If not, contact authorized service personnel. After 15
minutes, all patient information and trend data is lost and the monitor returns to the
user default settings (start-up mode). For more information, see the technical manual.
WARNING
The B450 must always be used with a battery inserted. This
will ensure the functioning of the monitor and prevent data
loss during possible supply mains interruptions.
Power failure alarm
B650 only.
If there is a power failure (e.g., the supply mains is interrupted and the monitor battery
fails), the monitor gives a continuous beeping alarm. This alarm remains active for as
long as there is some residual power left, or until it is silenced with the pause audio
alarm key, or until the power is reconnected.
106
CARESCAPE Modular Monitors
2062971-001
Summary of Contents for CARESCAPE
Page 38: ...38 CARESCAPE Modular Monitors 2062971 001 ...
Page 114: ...Setting up the monitor before use 114 CARESCAPE Modular Monitors 2062971 001 ...
Page 146: ...Alarms 146 CARESCAPE Modular Monitors 2062971 001 ...
Page 218: ...Pulse oximetry 218 CARESCAPE Modular Monitors 2062971 001 ...
Page 234: ...Non invasive blood pressure 234 CARESCAPE Modular Monitors 2062971 001 ...
Page 260: ...Temperature 260 CARESCAPE Modular Monitors 2062971 001 ...
Page 274: ...Cardiac output 274 CARESCAPE Modular Monitors 2062971 001 ...
Page 280: ...Mixed venous oxygen saturation SvO 280 CARESCAPE Modular Monitors 2062971 001 ...
Page 338: ...Patient Spirometry 338 CARESCAPE Modular Monitors 2062971 001 ...
Page 372: ...Neuromuscular transmission 372 CARESCAPE Modular Monitors 2062971 001 ...
Page 404: ...Laboratory data 404 CARESCAPE Modular Monitors 2062971 001 ...
Page 410: ...Calculations 410 CARESCAPE Modular Monitors 2062971 001 ...
Page 416: ...Drug calculations 416 CARESCAPE Modular Monitors 2062971 001 ...
Page 424: ...Trends 424 CARESCAPE Modular Monitors 2062971 001 ...
Page 432: ...Snapshots and events 432 CARESCAPE Modular Monitors 2062971 001 ...
Page 462: ...Cleaning and care 462 CARESCAPE Modular Monitors 2062971 001 ...
Page 528: ...Abbreviations yr year yrs years 528 CARESCAPE Modular Monitors 2062971 001 ...
Page 546: ...Skills checklist 546 CARESCAPE Modular Monitors 2062971 001 ...
Page 547: ...content ...