CARESCAPE V100 Vital Signs Monitor
3-5
Installation:
Configuring the monitor
Configuring the monitor
Operating modes
The monitor can operate in one of four modes: clinical, configuration, advanced
configuration, and service.
Clinical mode
Clinical mode is the monitor’s normal operating mode. While this mode is active,
alarm limits and a few other commonly used settings are adjustable. All
parameters are available for monitoring in this mode.
Configuration mode
Configuration and advanced configuration modes display the software revision
and allow you to configure defaults for some settings that are available in
clinical mode, as well as some less commonly used settings that are only
adjustable in these modes.
A fatal error history is also available in the advanced
configuration mode. No parameters are operable in these modes, therefore,
patient monitoring is suspended.
Configuration mode settings
Monitor settings such as
HIGH
/
LOW
alarm settings changed in the Clinical Mode
will not be retained after the monitor is powered off. To retain alarm and
parameter settings, the changes must be done in the configuration mode. Date/
Time settings are also entered in the configuration mode.
To enter the configuration mode: with the monitor off, press and hold the
Menu
button at the same time as pressing and holding the
On/Off
button for 3
seconds. The monitor enters the configuration mode.
For a few seconds immediately after power up in this mode, the
Systolic
and
Diastolic
windows display the major and minor version codes. The version codes
are numbers that represent the letters of the English alphabet, which are
designated to the currently loaded version of the monitor firmware (e.g., 1
indicates A, 2 indicates B, etc.).
At the same time, the NIBP Algorithm selected in the monitor is displayed in the
min
(minutes display) window as follows:
AUSC
if the monitor is configured with auscultatory NIBP Algorithm
STAT
if the monitor is configured with DINAMAP SuperSTAT Algorithm
CLAS
if the monitor is configured with DINAMAP Classic Algorithm
Also, the Temperature probe selected in the monitor is displayed in the
Temperature
window as follows:
trb0
if the monitor is configured for Alaris Turbo Temp
trI
if the monitor is configured for Alaris Tri-Site
tat
if the monitor is configured for Exergen
Summary of Contents for CareScape V100
Page 9: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 1 1 1 Introduction ...
Page 25: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 2 1 2 Equipment overview ...
Page 43: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 3 1 3 Installation ...
Page 59: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 4 1 4 Maintenance ...
Page 90: ...4 32 CARESCAPE V100 Vital Signs Monitor 2048723 002C Maintenance Test results form ...
Page 91: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 5 1 5 Troubleshooting ...
Page 99: ...2048723 002C CARESCAPE V100 Vital Signs Monitor 6 1 6 Parts lists and drawings ...
Page 163: ...2048723 002C CARESCAPE V100 Vital Signs Monitor B 1 B Appropriate use of NIBP simulators ...
Page 177: ......