The S/5 virtual ID change will be applied immediately. All other network configurations will be
saved and become active when the patient monitor is restarted
7.2.4 Configuring ADT server settings
1.
Log in to Webmin.
2.
Select
Configuration
>
Network
>
ADT settings
.
3.
In the
ADT settings
window, select
Enable
to enable the Visit Number Query.
4.
Enter the ADT IX IP address.
7.2.2 Selecting and configuring CARESCAPE Network
for a list of reserved
addresses that must not be used as the ADT IX IP address.
5.
Select
Save
.
To disable the Visit Number Query, select
Disable
in the
ADT settings
window.
7.3 Setting time and date
NOTE: In the S/5 Network, the patient monitor’s date and time is automatically synchronized
with the date and time of the iCentral it is connected to. You can set the date and time locally
only if the patient monitor is not connected to the S/5 Network and the patient monitor is in a
discharged state.
CAUTION
NETWORK DEVICE TIME SYNCHRONIZATION — When adding a new device to
the CARESCAPE Network, the existing devices on the CARESCAPE Network will
synchronize to the new device’s time. To prevent potential time
synchronization issues, you should set the new device’s time to be as close as
possible to the time (within one minute or less) used by the existing GE devices
on the CARESCAPE Network.
1.
Log in to Webmin.
2.
Select
Configuration
>
Time
.
3.
In the
Time Configuration
window, in
Configure Date and Time
, update the following
fields as needed:
•
Date
•
Month
•
Year
•
Hour:Minute
•
AM/PM
•
12/24 Hrs
4.
In the
Time Configuration
window, in
Configure Time Zone
, select the appropriate
UTC
Offset
, if applicable.
NOTE: The time zone configuration only applies to communication with Citrix server.
5.
Select
Save
.
The manual time configuration takes effect immediately.
7.4 Setting unit and bed name
Configure the care unit name and bed name for patient monitors that are configured to
connect to the MC Network.
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......