Pre-installation requirements
51
2106778-001
5.4 Installing the mounting hardware
Ensure that all the applicable/ required mounting hardware is properly installed prior to the
installation of the patient monitor:
•
Mounting hardware for the patient monitor, either for a stand-alone installation or for an
installation to an anesthesia machine or to a ventilator
•
Mounting hardware for the module frame
•
Mounting hardware for the PSM module
•
Mounting hardware for the PDM module
•
Mounting hardware for the displays
•
Mounting hardware for the Unity Network ID connectivity device
NOTE: Refer to the patient monitor’s supplemental information manual for compatible
mounting hardware for each system component above. Each mounting kit includes the
necessary hardware and the installation instructions.
WARNING
Use only manufacturer specified mounts.
5.5 Unity Network ID connectivity device installation
The Unity Network ID connectivity device shall be properly installed, configured and tested
according to the Unity Network ID Connectivity Device Service Manual prior to connecting it to
the patient monitor.
Make sure that the Unity Network ID connectivity device is configured as follows:
•
IP address is 192.168.253.x, where x is a number between 2 and 254.
•
Netmask is 255.255.255.0
•
The location of the Unity Network ID is set to a value other than the default (XXXX-XXX).
For example, BAY3|UNID3+.
Refer to the Unity Network Interface Device (ID) Service Manual for instructions on checking
and changing the IP address.
5.6 Power and environmental requirements
Check the patient monitor’s supplemental information manual for power and environmental
requirements.
WARNING
Operation of the monitor outside the specified performance range may
cause inaccurate results.
CAUTION
Do not use or store equipment outside the specified temperature, humidity, or
altitude ranges.
Power requirements
•
The installation site shall have hospital-grade grounded power outlets and power cords
for all system components.
•
Verify that the power outlet is wired correctly according to the country’s electrical code
standard.
WARNING
EXCESSIVE LEAKAGE CURRENT - A display or printer that is a non-medical
grade device and is used within the patient environment, must always be
powered from an additional transformer providing at least basic isolation
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......