CARESCAPE Monitor B850
128
2106778-001
4.
Verify that the recorder starts printing. Let the recorder print for approximately 10
seconds and verify the following things from the printout:
The header line contains the date, time and some other applicable status and
configuration information.
The grid is clear.
The waveforms labels appear in the printout as configured.
Select one of the following options to stop printing:
Select
Stop Printing
from the main screen.
Select
Monitor Setup > Printing > Waveforms > Stop Printing
.
10.3.13 Test completion
Select
Discharge Patient
or
Reset Case
to discard any changes made to the patient monitor
configuration during checkout.
•
Complete the Appendix
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......