8838 User Guide (BB1857-H)
The transducer is connected to the system using both the array Transducer Socket and
the mechanical Transducer Socket.
To connect the array Transducer Socket
1
Insert transducer plug into socket with locking lever in 3 o’clock position.
2
Turn locking lever on transducer plug clockwise to 6 o’clock position.
To connect the mechanical Transducer Socket
1
Align red mark on plug with red dot on transducer socket.
2
Insert plug in socket.
When connected, the transducer complies with Type B requirements of EN60601-1
(IEC 60601-1).
Changing Frequency
The Multi-Frequency Imaging (MFI) facility enables you to select the imaging
frequency. See the applicable system user guide for instructions. The selected
frequency is displayed at the top of the screen.
Using a Transducer Cover
You must use a transducer cover. See the Product Data sheet for a list of available
transducer covers, including sterile covers.
Apply imaging gel or other water-soluble agent inside and outside the cover to create
good acoustic contact.
To put on the cover:
1
Apply gel to the tip of the transducer so that it covers the entire imaging surface
(most of the black part of the transducer). You can also put imaging gel inside
the tip of the cover before you put the cover over the transducer.
Caution
BK-c4
The system must not be imaging when you connect the transducer. The image must be
frozen or the system must be turned off. Do not unfreeze the image before you have
connected the transducer to
both
sockets. Otherwise the imaging array in the
transducer can be parked in an incorrect position, which can lead to incorrect 3D
representations of the regions being imaged.
WARNING
BK-w13
Do not connect two 8838 transducers simultaneously. The system will produce an
incorrect 3D volume with only one identical image repeated throughout the entire
volume.
WARNING
TC-w1
Some transducer covers can contain latex. Because of reports of severe allergic reactions to
medical devices containing latex (natural rubber), the FDA advises health-care professionals
to identify their latex-sensitive patients and be prepared to treat allergic reactions promptly.