The transducer is connected to the system using both the array Transducer Socket and 
the mechanical Transducer Socket. 

To connect the array Transducer Socket

1

Insert transducer plug into socket with locking lever in 3 o’clock position.

2

Turn locking lever on transducer plug clockwise to 6 o’clock position. 

To connect the mechanical Transducer Socket

1

Align red mark on plug with red dot on transducer socket.

2

Insert plug in socket.

When connected, the transducer complies with Type B requirements of EN60601-1 
(IEC 60601-1).

Changing Frequency

The Multi-Frequency Imaging (MFI) facility enables you to select the imaging 
frequency. See the applicable system user guide for instructions. The selected 
frequency is displayed at the top of the screen.

Using a Transducer Cover

You must use a transducer cover. See the Product Data sheet for a list of available 
transducer covers, including sterile covers.

Apply imaging gel or other water-soluble agent inside and outside the cover to create 
good acoustic contact.

To put on the cover:

1

Apply gel to the tip of the transducer so that it covers the entire imaging surface 
(most of the black part of the transducer). You can also put imaging gel inside 
the tip of the cover before you put the cover over the transducer.

Caution 

BK-c4

The system must not be imaging when you connect the transducer. The image must be 
frozen or the system must be turned off. Do not unfreeze the image before you have 
connected the transducer to 

both

 sockets. Otherwise the imaging array in the 

transducer can be parked in an incorrect position, which can lead to incorrect 3D 
representations of the regions being imaged.

WARNING

BK-w13

Do not connect two 8838 transducers simultaneously. The system will produce an 
incorrect 3D volume with only one identical image repeated throughout the entire 
volume. 

WARNING

TC-w1

Some transducer covers can contain latex. Because of reports of severe allergic reactions to 
medical devices containing latex (natural rubber), the FDA advises health-care professionals 
to identify their latex-sensitive patients and be prepared to treat allergic reactions promptly.

Summary of Contents for BK MEDICAL 8838 Series

Page 1: ...For Professional Users Only English BB1857 H April 2023 User Guide Type 8838 3DART Transducer...

Page 2: ...products and services As part of our customer satisfaction program we contact a sample of our customers a few months after they receive their orders If you receive an email message from us asking for...

Page 3: ...ing and Disinfection 7 Starting Imaging 7 Connecting the Transducer 7 Changing Frequency 8 Using a Transducer Cover 8 Using the Transducer Control Button 9 Changing Orientation 9 Endovaginal Anorectal...

Page 4: ...4...

Page 5: ...suitable for contrast imaging1 Patient Population The patient population is adults Figure 1 3DART Transducer Type 8838 Imaging Plane The transducer contains a linear array of ultrasound elements that...

Page 6: ...with the patient Turn off the system Unplug the system from the wall and make sure it cannot be used until it has been checked Do not try to repair the system yourself Contact your BK service represen...

Page 7: ...n and responsibility to provide the highest possible degree of infection control to patients co workers and themselves The instructions in this book are meant as a guide To avoid cross contamination f...

Page 8: ...oluble agent inside and outside the cover to create good acoustic contact To put on the cover 1 Apply gel to the tip of the transducer so that it covers the entire imaging surface most of the black pa...

Page 9: ...e the orientation of the image on the monitor refer to the applicable system user guide for instructions To change the angle of the 2D imaging plane see Rotating the 2D Imaging Plane on page 10 Endova...

Page 10: ...the position of the transducer array Note that the default position of the imaging plane is opposite the transducer button see Figure 2 Figure 4 Rotation symbol on the flex Focus 1202 Not available o...

Page 11: ...to its default position You should always check the position of the imaging plane to make sure that it is in the desired postition 3D Imaging NOTE See the advanced system user guide for more informat...

Page 12: ...the stitch angle is opposite the array s default position that is towards the transducer button see Figure 6 Figure 6 The stitch angle s default position 2 Select the position you want by choosing the...

Page 13: ...change this parameter on the system Untracked 3D acquisition available with some transducers is not an option for this transducer Select the Mover parameter on the screen There are two options IP8838...

Page 14: ...lowed Within the EU when you discard the transducer you must send it to appropriate facilities for recovery and recycling See the applicable system user guide for further details WARNING D w1 For disp...

Page 15: ......

Page 16: ......

Reviews: