2062665-001 G

 

PCD-01: HIS Disconnection or Network Outage

4

PCD-01

Summary of Contents for B40

Page 1: ...GE Healthcare B40 B20 Patient Monitor HL7 Reference Manual B40 B20 Patient Monitor English 2062665 001 G paper 2014 General Electric Company All Rights Reserved...

Page 2: ...o change without notice NOTE For technical documentation purposes the abbreviation GE is used for the legal entity name GE Healthcare Finland Oy Listed below are GE Medical Systems Information Technol...

Page 3: ...ge signal words 4 2 HL7 Interface 5 HL7 Standard Background 6 Overview 6 Multiple Segments 6 Data Elements 6 HL7 Communication 7 Observation results interfaces 7 3 Observation results interface 8 Gene...

Page 4: ...onitor HL7 interface 18 Result reporting messages ORU 18 Overview 18 Result Message Composition 18 Result Message Examples 19 Application High Level Acknowledgment Messages 24 Overview 24 Acknowledgme...

Page 5: ...2062665 001 G 3 B Abbreviations 45 Abbreviations 46...

Page 6: ...4 2062665 001 G...

Page 7: ...2062665 001 G 1 Introduction 1 Introduction...

Page 8: ...to provide instructions for implementing and using the HL7 standard Details of the HL7 standard can be found in Health Level Seven Version 2 6 Contains data formats for the transmission of data and de...

Page 9: ...horized service personnel only Any unauthorized attempt to repair equipment under warranty voids the warranty For GE s products it is the user s responsibility to report the need for service to GE or...

Page 10: ...y CAUTION indicates a hazardous situation that if not avoided could result in minor or moderate injury NOTICE indicates a hazardous situation not related to personal injury that if not avoided could r...

Page 11: ...2062665 001 G 5 HL7 Interface Service Information 2 HL7 Interface...

Page 12: ...varying lengths Like the segments they are separated from each other by special separator characters Certain data elements and their separators are logically grouped together to create a data segment...

Page 13: ...ng items message formats acknowledgment protocol communication protocol data handling Communication between HIS personnel and GE interface personnel is essential to determine the customer specific use...

Page 14: ...8 2062665 001 G Observation results interface HL7 Standard Background 3 Observation results interface...

Page 15: ...terface the monitor can return result messages to the HIS The Patient Monitor Outbound implementation of the HL7 outbound interface protocol provides result messages The HL7 result reporting interface...

Page 16: ...he GE HL7 interface uses TCP IP sockets for all outbound messages and conforms to the IEEE 802 3 Ethernet standard The HL7 interface assumes that the low level communication protocol ensures that the...

Page 17: ...blished The monitor continuously tries to connect to the HIS on a regular periodic interval using the auto connection feature The monitor transfers the HL7 messages to the HIS server by TCP IP socket...

Page 18: ...12 2062665 001 G PCD 01 HIS Disconnection or Network Outage 4 PCD 01...

Page 19: ...device in OBX 18 equipment instance identifier The value of this identifier is composed of two components Component 1 is an EUI 64 identifier that starts with the GE Healthcare company code plus the...

Page 20: ...ram Physiological Monitor Containment tree Physiological monitors are comprised of a number of different Virtual Medical Devices VMDs as indicated in the following example containment tree The first O...

Page 21: ...er and then the discrete observations OBX 2 69854 MDC_DEV_METER_PRESS_BLD_VMD MDC 1 13 0 0 X OBX 3 69855 MDC_DEV_METER_PRESS_BLD_CHAN MDC 1 13 1 0 X OBX 4 NM 150033 MDC_PRESS_BLD_ART_SYS MDC 1 13 1 1...

Page 22: ...666 1 6 0 0 Gas Mon MDC_DEV_GEN_CONC_AWAY_VMD 69766 1 11 0 0 IP MDC_DEV_METER_PRESS_BLD_VMD 69854 1 13 0 0 IP MDC_DEV_METER_PRESS_BLD_CHAN 69855 1 13 x 0 where x is the slot number NIBP MDC_DEV_PRESS_...

Page 23: ...2062665 001 G 17 Sample Messages PCD 01 description 5 Sample Messages...

Page 24: ...of monitor is mapped to data field positions in the HL7 segment structures to create result messages The following pages provide the general format of the various HL7 data messages recognized by the m...

Page 25: ...essages The monitor HL7 outbound interface can generate countless variations of these examples Example 1 This example has one OBR segment followed by the corresponding OBX segments collected at a user...

Page 26: ...NM 150087 MDC_PRESS_BLD_VEN_CENT_MEAN MDC 1 13 2 1 1 266016 MDC_DIM_MMHG MDC R 080019FFFE1 34535 B40_GE OBX 10 69798 MDC_DEV_ECG_VMD MDC 1 5 0 0 X OBX 11 NM 147842 MDC_ECG_HEART_RATE MDC 1 5 1 1 21 2...

Page 27: ...16 MDC_DIM_MMHG MDC R 080019FFFE0B4020 B40_GE OBX 6 NM 150035 MDC_PRESS_BLD_ART_MEAN MDC 1 13 1 3 9 5 266016 MDC_DIM_MMHG MDC R 080019FFFE0B402 0 B40_GE OBX 7 NM 149522 MDC_BLD_PULS_RATE_INV MDC 1 13...

Page 28: ...66419 MDC_DIM_MICRO_VOLT MDC R 080019FFFE0 B4020 B40_GE OBX 20 NM 131847 MDC_ECG_AMPL_ST_V5 MDC 1 5 1 10 0 00 266419 MDC_DIM_MICRO_VOLT MDC R 080019FFFE 0B4020 B40_GE OBX 21 NM 131848 MDC_ECG_AMPL_ST_...

Page 29: ...0019FFFE0B4020 B40_GE OBX 34 NM 0 MDC_CONC_MAC_SUM_AGE_CORR MDC 1 11 1 9 0 262656 MDC_DIM_DIMLESS MDC R 080019FFFE0B40 20 B40_GE OBX 35 NM 152312 MDC_AWAY_ISOFL_EXP MDC 1 11 1 10 0 2 62688 MDC_DIM_PER...

Page 30: ...ssages Overview The monitor Outbound HL7 interface supports application level message acknowledgments formatted as HL7 original mode acknowledgments This function allows the HIS receiving application...

Page 31: ...l attempts AR messages are also not retransmitted the system will disconnect the connection with the HL7 receiving system If the acknowledgment configuration is enabled and no HL7 Acknowledgement mess...

Page 32: ...nitor HL7 interface MSA AR 17f1be29 aa2a 11d2 a655 00105a1e9b67 Unsupported message type Acknowledgment Error MSH NUR MMS HL7 199812021002 ACK R01 17f1be29 aa2a 11d2 a655 00105a1e9b67 P 2 3 MSA AE 17f...

Page 33: ...2062665 001 G 27 HL7 Data Segment Definitions Transactions from the monitor HL7 interface 6 HL7 Data Segment Definitions...

Page 34: ...onitoring devices The Y simply indicates that the field is supported and would not be ignored as the fields marked N should it be received However the HL7 application is strictly outbound at this poin...

Page 35: ...ite ID w check digit TS Time stamp DLN Driver s license number TX Text data DT Date XAD Extended composite ID number HD Hierarchic designator name for persons ID Coded value XON Extended composite nam...

Page 36: ...eceiving Facility 7 26 DTM R Date Time of Message 8 40 ST RE Security 9 15 MSG R Message Type 10 20 ST R Message Control ID 11 3 PT R Processing ID 12 60 VID R Version ID 13 15 NM RE Sequence Number 1...

Page 37: ...st 4 20 CX RE 00107 Alternate Patient ID PID 5 250 XPN RE 00108 Patient Name 6 250 XPN RE 00109 Mother s Maiden Name 7 26 TS X 00110 Date Time Birth 8 1 IS RE 0001 00111 Administrative Sex 9 250 XPN X...

Page 38: ...8 250 CE RE 0212 00739 Nationality 29 26 TS RE 00740 Patient Death Data and Time 30 1 ID RE 0136 00741 Patient Death Indicator 31 1 ID RE 0136 01535 Identity Unknown Indicator 32 20 IS RE 0445 01536 I...

Page 39: ...Service 11 80 PL X Temporary Location 12 2 IS X 0087 Preadmit Test Indicator 13 2 IS X 0092 Re admission Indicator 14 6 IS X 0023 Admit Source 15 2 IS X 0009 Ambulatory Status 16 2 IS X 0099 VIP Indi...

Page 40: ...ition 37 47 DLD X 0113 Discharged to Location 38 250 CE X 0114 Diet Type 39 2 IS X 0115 Servicing Facility 40 1 IS X 0116 Bed Status 41 2 IS X 0117 Account Status 42 80 PL X Pending Location 43 80 PL...

Page 41: ...R Set Id 2 2 ID RE 0125 Value Type 3 250 CE R Observation Identifier 4 20 ST R Observation Sub ID 5 99999 Varies RE Observation Value 6 250 CE RE Units 7 60 ST RE Reference Range 8 5 IS RE 0078 Abnor...

Page 42: ...36 2062665 001 G HL7 Data Segment Definitions Data Segments 19 26 DTM CE Date Time of the Analysis 20 705 CWE RE 0163 Observation Site SEQ LEN DT USAGE TABLE ELEMENT NAME...

Page 43: ...2062665 001 G 37 HL7 Data Segment Definitions Data Segments A Supported parameters...

Page 44: ...266016 Pulmonary Artery Pressure Diastolic PA D mm Hg MDC_PRESS_BLD_ART_ PULM_DIA 150046 MDC_DIM_MMH G 266016 Pulmonary Artery Pressure Mean PA M mm Hg MDC_PRESS_BLD_ART_ PULM_MEAN 150047 MDC_DIM_MMH...

Page 45: ...lse Rate ART Pulse Rate beat min MDC_BLD_PULS_RATE_ INV 149522 MDC_DIM_BEAT _PER_MIN 264864 UAC Pulse Rate UAC Pulse Rate beat min MDC_BLD_PULS_RATE_ INV 149522 MDC_DIM_BEAT _PER_MIN 264864 Right Vent...

Page 46: ...ure Mean LAP M mm Hg MDC_PRESS_BLD_ATR_ LEFT_MEAN 150067 MDC_DIM_MMH G 266016 Umbilical Arterial Catheter Pressure Systolic UAC S mm Hg MDC_PRESS_BLD_ART_ UMB_SYS 150057 MDC_DIM_MMH G 266016 Umbilical...

Page 47: ...DIM_PERC ENT 262688 Expired N2O N2O EX MDC_AWAY_N2O_EXP 152320 MDC_DIM_PERC ENT 262688 Inspired N2O N2O IN MDC_CONC_AWAY_N2 O_INSP 152192 MDC_DIM_PERC ENT 262688 MAC Sum MAC_SU M unitles s 1 MDC_CONC_...

Page 48: ...DC_DIM_MICR O_VOLT 266419 ST Segment III ST III uV MDC_ECG_AMPL_ST_III 131901 MDC_DIM_MICR O_VOLT 266419 ST Segment V1 ST V1 uV MDC_ECG_AMPL_ST_V1 131843 MDC_DIM_MICR O_VOLT 266419 ST Segment V2 ST V2...

Page 49: ...SAT_ O2 150456 MDC_DIM_PERC ENT 262688 Respirati on RR Respirati on Rate Respiration Rate RR breath min MDC_RESP_RATE 151562 MDC_DIM_RESP _PER_MIN 264928 Tempera ture TP Temperature 1 TP 1 Cel MDC_TEM...

Page 50: ...C_DIM_DEGC 268192 Core Temperature TP Core Cel MDC_TEMP_CORE 150368 MDC_DIM_DEGC 268192 Surface Temperature TP Surf Cel MDC_TEMP_BODY 150364 MDC_DIM_DEGC 268192 Entropy Suppression Ratio BURST SUPPRE...

Page 51: ...2062665 001 G 45 HL7 Data Segment Definitions Supported parameters B Abbreviations...

Page 52: ...ormation Interchange ADD Addendum C CCG Centricity Clinical Gateway CE coded entry CHAR character Chg change CIS Clinical Information System CK Composite ID with check digit CM composite CN composite...

Page 53: ...ervation request Obs observation OBX observation result O optional P PDM Patient Data Model PN part number person name PID Patient Identification PV1 Patient Visit Q QRD Query Definition R Req request...

Page 54: ...48 2062665 001 G HL7 Data Segment Definitions Abbreviations...

Page 55: ......

Page 56: ...ormation Technologies Inc 8200 West Tower Avenue Milwaukee WI 53223 USA Tel 1 414 355 5000 1 800 558 5120 US only Fax 1 414 355 3790 Asian Headquarters GE Medical Systems Information Technologies Asia...