Alarms
4-11
Deactivating alarms
You can set the alarm limits temporarily to OFF for the following parameters: HR, SpO
2
and
respiration rate. This way, you can silence a parameter without having to adjust the alarm
limits.
1.
Press the
Alarms Setup
key.
2.
Select
Adjust Limits
.
3.
Turn the ComWheel to highlight the measurement.
4.
and push the ComWheel to change the selection
Off
.
The symbol
appears in the digit field.
When you select the alarm limit ON again, the previous alarm limits will be active.
Automatic recording on alarms
An automatic local recording is possible when Asystole, Tachy/Brady, Art High/Low, V Fib or V
Tachy alarms reach the red alarm level. When the red alarm level is reached, the recorder
prints the ECG + Art waveforms.
1.
Press the
Print/Record
key.
2.
Select
Record Waveforms
.
3.
Select
Start on Alarms
and
Yes
.
Print/
Record
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......