Alarms
4-5
Alarms Setup menu
You can view and adjust patient alarm limits in the
Alarms Setup
menu.
(1) List
of
selections
(2)
Exit
from the alarm limit adjustment area back to
Alarms Setup
menu
(3) Parameter box with high and low limit values and a 10-minute trend showing the current
status
Adjust Limits
Adjusts individual measurement alarm limits. You can also access the adjustment menu
through each parameter menu.
NOTE: If the monitor is connected to the network, the alarm limits can also be changed
using the Central if this feature has been enabled in its configuration.
Auto Limits
Auto limits are calculated from the displayed patient reading at the time when auto limits are
selected.
Default Limits
Sets the alarms to the default alarm limits.
NOTE: Default limits can be changed and saved to modes. For more information, see the
"Default Configuration Worksheet."
Cancel Changes
Returns all the limits to the ones set before entering the
Alarms Setup
menu if you have not
exited the menu yet.
Arrh. Alarms
Opens a menu to select and adjust
Severe
arrhythmia analysis. Alarm priorities can be
configured in this menu.
Adjust
= defines the alarm priorities (red, yellow, white, off).
NOTE: You can only set the priority for V Tachy.
NOTE: V Tachy alarm cannot be selected OFF.
NOTE: Alarm priorities can also be set using the Central, depending on its
configuration.
To redisplay the
Alarms Setup
menu, select
Alarms Setup
.
For more information on arrhythmia alarms, please see section
Alarms
Setup
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......