Invasive blood pressure
14-7
Table 14-1
Invasive blood pressure labels and descriptions
Cerebral perfusion pressure
Cerebral Perfusion pressure CPP is calculated by subtracting ICP mean pressure from Art mean
pressure.
CPP mean is displayed next to the ICP value.
NOTE: Set the
Label
of the channel to
ICP
, and the
Digit Format
to
CPP
.
Adjusting alarm sources and limits
You can adjust or turn off pressure alarm limits in the
Alarms Setup/Adjust Limits
menu. Enter
the menu by pressing the
Alarms Setup
key.
For each pressure channel you can choose as a source
Sys
(Systolic),
Dia
(Diastolic),
Mean
or
Off
. Note that you can choose one or both sources to be active at a time.
For more information about alarms and adjusting, see section
LABEL
P1, Art,
ABP
P2, CVP RAP, LAP
ICP
PA
RVP
Scale mmHg/kPa
200/30
20/3
20/3
20/3
60/8
60/8
Color
Red
Blue
White
White
Yellow
White
Alarm source
Sys
Mean
Off
Off
Off
Off
Digit format
S/D
Mean
Mean
CPP
S/D
S/D
Filter (Hz)
22
9
9
9
9
9
Response
normal
normal
normal
normal
normal
normal
Alarms
Setup
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......