Invasive blood pressure
14-1
14 Invasive blood pressure
Overview
You can measure and monitor two invasive blood pressures at the same time using a dual
invasive blood pressure cable. To measure invasive blood pressure you need the E-PSMPW
module.
During the invasive blood pressure measurement, the transducer converts pressure variations
into electrical signals. The electrical signals are amplified and displayed as numeric pressure
values and waveforms.
WARNING
All invasive procedures involve risks to the patient. Use aseptic technique.
Follow catheter manufacturer’s instructions.
Module description
Module keys
E-PSMPW module has direct function keys to zero both pressure transducers separately.
•
Press the
Zero
P1
or
Zero P2
module key. The message ‘Zeroing’ is displayed. After the
transducer is zeroed, the message ‘Zeroed’ is displayed.
Side panel key
The monitor has one side panel key for invasive pressures, the
Zero ALL
key. You can use this
to zero all pressure transducers.
NOTE: Selecting
Zero ALL
does not zero ICP. Zero it separately.
E-PSMPW
Invasive pres-
sures connector
Summary of Contents for b30
Page 2: ......
Page 8: ......
Page 10: ...Table of contents ii...
Page 12: ...Table of contents ii...
Page 16: ...B30 Patient Monitor 4...
Page 18: ...Table of contents ii...
Page 26: ...Table of contents ii...
Page 62: ...B30 Patient Monitor 2 36...
Page 64: ...Table of contents ii...
Page 72: ...Table of contents ii...
Page 90: ...Table of contents ii...
Page 106: ...Table of contents ii...
Page 118: ...B30 Patient Monitor 6 12...
Page 120: ...Table of contents ii...
Page 124: ...B30 Patient Monitor 7 4...
Page 126: ...Table of contents ii...
Page 136: ...Table of contents ii...
Page 144: ...B30 Patient Monitor 9 8...
Page 146: ...Table of contents ii...
Page 156: ...Table of contents ii...
Page 176: ...Table of contents ii...
Page 184: ...B30 Patient Monitor 12 8...
Page 186: ...Table of contents ii...
Page 192: ...Table of contents ii...
Page 202: ...Table of contents ii...
Page 212: ...B30 Patient Monitor 15 10...
Page 214: ...Table of contents ii...
Page 224: ...B30 Patient Monitor 16 10...
Page 226: ...Table of contents ii...
Page 236: ...B30 Patient Monitor 17 10...
Page 241: ......