![GE b30 Technical Reference Manual Download Page 204](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170204.webp)
B30 Patient Monitor
8-4
Document no
.
2044677-001
2.2.4 Conditions exceeding normal
Accuracy specifications under the following conditions
nop
:
n
Ambient temperature
10...40
°
C, within
±
5
°
C of calibration
Ambient pressure
500...800 mmHg,
±
50 mmHg of calibration
Ambient humidity
10...98% RH,
±
20% RH of calibration
o
During warm-up 1 to 10 minutes, under normal conditions
p
During warm-up 10 to 30 minutes, under normal conditions
2.3 Respiration Rate (RR)
Measurement range
4 to 80 breaths /min
Breath detection
1% change in CO
2
level
Accuracy
±1/
min in the range 4 to 20 l/min
±5% in the range 20 to 80 l/min
Resolution
1/min
Adjustable low and high limits for respiration rate; alarm for apnea.
2.4 Recorder
2.4.1 Typical performance
Principle
Thermal array
Print resolution
Vertical
8 dots/mm (200 dots/inch)
Horizontal
32
dots/mm (800 dots/inch) at a speed of 25 mm/s and slower
Paper width
50 mm, printing width 48 mm
Print speed
1, 6.25, 12.5, 25 mm/s
Waveform printout
selectable 1, 2, or 3 waveforms
Numerical printout
HR, NIPBP, P1, P2, T1, T2, Tblood, Et/FiCO
2
, RR
Tabular Trend printout
HR, SpO
2
, P1, NIBP, EtCO
2
Accuracy under different conditions (see above)
Conditions
n
and
p
Condition
o
CO
2
(0 … 15 vol%)
±
(0.3 vol% + 4% of reading)
(at 5 vol% error
±
0.5 vol%)
±
(0.4 vol% + 7% of reading)
(at 5 vol% error
±
0.75 vol%)
CO
2
(15 … 20 vol%)
±
(0.8 vol% + 4% of reading)
(at 5 vol% error
±
0.5 vol%)
±
(0.9 vol% + 7% of reading)
(at 5 vol% error
±
0.75 vol%)
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......