Specifications
8-3
Document no. 2044677-001
2.2.2 Technical specifications
Airway humidity
0...100%, condensing
Sampling rate
150 ±25ml/min (sampling line 2 to 3 m, normal conditions)
Sampling delay
2.1 seconds typical with a 3-m sampling line
Total system response time
2.4 seconds typical with a 3-m sampling line, including sampling
delay and rise time (typically 3.7 seconds with a 6-m sampling
line).
Automatic compensation for barometric pressure, CO
2
-NO
2
and CO
2
-O
2
collision broadening effect
compensation selectable from menu.
Warm-up time
1 min for operation with CO
2,
30 min for full specification
Zeroing interval
4, 15, 30 and 60 minutes after start-up, then every 60 minutes.
2.2.3 Normal conditions
Accuracy specifications apply in normal conditions (after 30 minutes warm-up period):
Ambient temperature
18...28
°
C, within
±
5
°
C of calibration
Ambient pressure
500...800mmHg, ±50mmHg of cal.
Ambient humidity
20...80% RH, ±20% RH of cal.
Automatic compensation for barometric pressure.
Non-disturbing gases are those with a maximum effect on the CO
2
reading <0.2 vol%. The
effect is valid for specific concentrations shown in parentheses of the non-disturbing gas:
−
Ethanol C
2
H
5
OH (<0.3%)
−
Acetone (<0.1%)
−
Methane CH
4
(0.2%)
−
Nitrogen N
2
−
Water vapor
−
Dichlorofluromethane (<1%)
−
Tetrafluoroethane (<1%)
Disturbing gases and their effect on the CO
2
reading at 5.0 vol% CO
2
are shown below. Errors
listed reflect the effect of specific concentrations (shown in parentheses) of an individual
disturbing gas and should be combined when estimating the effect of gas mixtures:
−
Halotane (4%) increases < 0.3 vol%
−
Isoflurane (5%) increases < 0.4 vol%
−
Enflurane (5%) increases < 0.4 vol%
−
Desflurane (24%) increases < 1.2 vol%
−
Sevoflurane (6%) increases < 0.4 vol%
−
Helium (50%) decreases < 0.3 vol%
−
If O
2
compensation is not activated: O
2
(40 … 95%) decreases < 0.3 vol%
−
If O
2
compensation is activated: O
2
(40 … 95%) error < 0.15 vol%
−
If N
2
O compensation is not activated: N
2
O (40%) increases < 0.4 vol%
−
N
2
O (40 to 80%) increases < 0.8 vol%
−
If N
2
O compensation is activated: N
2
O (40 to 80%) error < 0.3 vol%
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......