E-PSM(P)W Specifications
7-3
Document no. 2044677-001
2
E-PSM(P)W Specifications
2.1 General specifications
Module size
55 x 135 (175 w/ tab) x 145 mm
W x D x H
2.2 x 5.3 (6.9 w/ tab) x 5.7 in
Module weight
0.6 kg /1.4 lb.
Operation temperature
5 to 40
°
C / 41 to 104
°
F
2.2 Typical performance
2.2.1 NIBP
NOTE: Non-invasive blood pressure measurement is intended for patients weighing over 5 kg
(11 lb.)
Oscillometric measurement principle.
Measurement range
adult
25 to 260 mmHg
child
25 to 190 mmHg
infant
15 to 140 mmHg
Pulse rate range accepted
30 to 250 bpm
Typical measuring time
adult
< 30 s
infant
< 25 s
Inflation limit
adult
280 ±10 mmHg
child
200 ±10 mmHg
infant
150 ±5 mmHg
Venous stasis
adult
80 ±5 mmHg / 2 min.
child
60 ±5 mmHg / 2 min.
infant
40 ±5 mmHg / 1 min.
Venous stasis pressure may be lower than the values above if the patient has low blood
pressure. The venous stasis pressure adapts to the measured mean pressure being the same
as mean pressure but always at least the following:
Infant 20 ± 5 mmHg
Child 30 ± 5 mmHg
Adult 40 ± 5 mmHg
2.2.2 ECG
Lead type
3-Lead (I, II, III)
5-Lead (I, II, III, aVR, aVL, aVF, V)
Sweep speeds
12.5, 25, 50 mm/sec.
Defibrillation protection
5000 V, 360 J
Recovery time
5 s
Input impedance
>2.5 M
Ω
(10 Hz)
Allowable offset
±1VDC
Gain range
0.2 to 5.0 cm/mV
Display filter
Diagnostic
0.05 to 150 Hz
Monitoring
0.5 to 30 Hz (-3 dB, with 50 Hz reject filter)
0.5 to 40 Hz (-3 dB, with 60 Hz reject filter)
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......