B30 Patient Monitor
6-12
Document no
.
2044677-001
3
Frame disassembly
WARNING
The B30 is always energized by the internal batteries. A short circuit may
cause internal damage. Do not touch any exposed wiring or conductive
surface inside, this may cause an electric shock.
CAUTION
Perform leakage current measurement whenever service or repair has been
done in the monitor.
NOTE: The backlight circuit runs on high voltage.
Do not touch the inverter board or the backlight tube leads when powered.
NOTE: Field repairs are recommended to the field replaceable unit (FRU) only. Attempting a field
repair on a factory sealed component or assembly could jeopardize the safe and effective
operation of the Monitor.
3.1 Before disassembly
NOTE: Wear a grounded, antistatic wristband when handling PC boards. Electrostatic
discharge may damage components on the board.
Make sure that the monitor is turned off.
Disconnect the main power cord. If the monitor is connected to the Datex-Ohmeda Network,
disconnect the RJ-45 cable.
NOTE: GE recommends that you assemble the monitor using the NEW fasteners (screws,
washers, etc.) provided in the FRU kits. Some fasteners, like the screws with a thread locking
coating, are NOT intended to be re-used more than three times.
Tools needed
- screwdrivers, TORX; T8, T10, T15
- crosshead screwdriver; M15
- flat blade screwdriver
- sockets; M4.5, M7, M11
- antistatic wristband
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......