B30 Patient Monitor
5-28
Document no
.
2044677-001
6
Service Log
Error, event and alarm data is stored in the Service
Log.
The service log contains information about the
occurred monitor errors, events and alarms since
the last factory reset or service log reset. The service
log is saved in the EEPROM memory of the main CPU
board.
Error Log
is for selecting the error history view onto
the right side of the menu. Error Log shows also
some monitoring events like warm and cold starts.
Alarm Log
is for selecting the alarm history view
onto the right side of the menu.
Scroll Log
is for scrolling the error/alarm
information on the right side of the menu.
Record Log
is for recording the service log
information onto the recorder.
Reset Log
is for clearing up the content of the
selected service log. This function should be run
after a performed maintenance.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......