Parameters
5-19
Document no. 2044677-001
4.3.1 Calibrations
How to calibrate T1/ T2
The calibrations are possible only when the protection is set
OFF
. The temperature calibration
requires accurate test plugs of value 25 °C and 45 °C.
1.
Select
Calibrate T1/Calibrate T2
2.
Insert the test plug 25 °C into the T1/T2 connector
3.
Press the ComWheel
4.
Insert the test plug 45 °C into the T1/T2 connector
5.
Press the ComWheel
How to calibrate P1/ P2
NOTE: Before starting the pressure calibration, disconnect all patient cables and discharge the
patient.
The calibrations require a pressure transducer (with appropriate cable) and a pressure
manometer.
1.
Connect the pressure transducer with the pressure manometer to the P1/P2 connector.
Select
Calibrate P1/Calibrate P2
. Leave the transducer to room air pressure.
2.
Press the ComWheel to start zeroing.
3.
Supply a pressure of 100 mmHg to 300 mmHg to the transducer. The recommended
pressure is 200 mmHg.
4.
Set the pressure on the display to match the pressure reading on the manometer and
press the ComWheel.
Protection
: Protection for the configuration and
temperature calibrations can be set ON and OFF.
Set Config
:
The module configuration should be set
according to the module type. The setting is
possible only when the protection is set OFF. The
available selections are TP, ST or STP. The
configuration setting should be checked if the STP
board is replaced.
Calibrate T1
/
Calibrate T2
:
The functions are for
calibrating the temperature channels T1 and T2.
Calibrate P1
/
Calibrate P2
:
The functions are for
calibrating the invasive blood pressure channels
P1 and P2.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......