135
Technical specifications
WARNING: Operation of the monitor outside the specified values may cause inaccurate results.
NOTE: Information in this section may be especially useful to clinicians.
General specifications
Size
B40 and B20 monitor
Without extension
module
312±5 mm (H) * 312±5 mm (W) * 158±5
mm (D)
With extension module
312±5 mm (H) * 352±5 mm (W) * 178±5
mm (D)
Weight
B40 with extension
module
≤
7 kg
B20 with extension
module
≤
6 kg
Monitor Environment
Operating temperature
Normal operation: 5 to 40°C (41 to
104°F)
Charging batteries: 5 to 35°C (41 to
95°F)
Storage and transport
temperature
-20 to 60°C (-4 to 140°F)
Operating humidity
20 to 90% noncondensing
Storage and transport
humidity
10 to 90% noncondensing
Operating atmospheric
pressure
700 to 1060 hPa
(525 to 795 mmHg)
Storage and transport
atmospheric pressure
700 to 1060 hPa
(525 to 795 mmHg)
E-miniC module Environment
Operating temperature
10 to 40°C (50 to 104°F)
Non-operating
temperature
-25 to 70°C (-13 to 158°F)
Operating humidity
10 to 95% noncondensing
Non-operating humidity
10 to 95% noncondensing
Operating altitude
666 to 1060 mbar
E-sCAiO, E-sCO, N-CAiO modules Environment
Operating temperature
10 to 40°C (50 to 104°F)
Non-operating
temperature
-25 to 60°C (-13 to 140°F)
Operating humidity
10 to 98% noncondensing
Non-operating humidity
10 to 90% noncondensing
Summary of Contents for B20
Page 2: ......
Page 18: ...14 E Entropy module 1 Module keys 2 Entropy connector 1 2 ...
Page 76: ...72 For your notes ...
Page 78: ...74 For your notes ...
Page 86: ...82 For your notes ...
Page 104: ...100 ...
Page 124: ...120 For your notes ...
Page 159: ......