Viewing alarm history
Use the
Alarm History
tab to view the list of the 12 most recent high
and medium priority alarms that occurred since the start of the case.
The alarm history clears at the start of a new case.
1.
Select
Alarm Setup
.
2.
Select the
Alarm History
tab.
The list of alarms shows in the window.
3.
Select
Close
.
Setting alarm volume
1.
Select
Alarm Setup
.
2.
Select the
Configure
tab.
3.
Set
Alarm Volume
to the desired value.
The alarm volume range is 1 to 5.
4.
Select
Close
.
Setting apnea delay
Use the
Apnea Delay
setting to set the desired apnea time delay.
The apnea time delay is the amount of time that can pass without the
system detecting a measured breath before the apnea alarm occurs.
1.
Select
Alarm Setup
.
2.
Select the
Configure
tab.
3.
Set
Apnea Delay
to the desired time.
•
The apnea time delay range is 10 to 30 seconds.
•
The apnea time settings are in 1 second increments.
4.
Select
Close
.
Silencing leak audio alarms
Use the
Leak Audio
setting to silence audio alarms of small leaks.
1.
Select
Alarm Setup
.
2.
Select the
Configure
tab.
3 Operation
2067226-001
3-19
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...