2.
To turn the volume apnea alarms off, select
Vol Apnea Alarm
to
Off
.
‘Volume Apnea Off’ shows in the general message field.
•
If mechanical ventilation is started, the volume apnea
alarms are active.
•
If manual ventilation is restarted, a pop-up confirmation
window appears to resume the
Off
setting.
3.
To turn the volume apnea alarms on, set
Vol Apnea Alarm
to
On
.
4.
Select
Close
.
Setting MV TV alarms
Use the
MV/TV Alarms
setting to turn off the MV and TV alarms.
Settings made during manual ventilation are not retained when
mechanical ventilation starts. Settings made during mechanical
ventilation are retained when manual ventilation starts.
For example, if
MV/TV Alarms
is set to
Off
during manual
ventilation, the alarms remain off until the Bag/Vent switch is set to
Vent or the
MV/TV Alarms
is set to
On
.
For example, if the
MV/TV Alarms
is set to
Off
during mechanical
ventilation, the alarms remain off when manual ventilation starts.
1.
Select
Alarm Setup
.
2.
To turn the volume alarms off, set
MV/TV Alarms
to
Off
.
‘MV/TV Alarms Off’ appears in the general message field.
The volume alarm limits waveform numerics shows dashes
during a case.
3.
To turn the volume alarms on, set
MV/TV Alarms
to
On
.
4.
Select
Close
.
Setting alarm limits
Do not set alarm limits to extreme values. Setting limits to
extreme values can render the alarm useless.
1.
Select
Alarm Setup
.
2.
From the
Primary Limits
and
More Limits
tabs, select the
alarm limit and make the change.
3.
Push the Home key, touch the waveform area of the display, or
select
Close
to close the menu.
WARNING
Aisys CS²
3-18
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...