Filling Aladin2 cassettes
Remove the cassette from the machine before filling the cassette.
Make sure that the cassette remains in a horizontal position during
filling. If the cassette is tilted, the flow of liquid is shut off to prevent
overfilling.
Make sure that the valve pins on the back of the cassette do not
come in contact with the system or any other object that could press
on the valve pins and let gas out of the cassette. Observe the liquid
level indicator of the cassette during filling. When the liquid level
reaches the full mark, stop filling.
To avoid spray from anesthetic agent after filling, slowly
remove the agent bottle from the filling port.
Always remove the cassette from the machine and put it
on a horizontal surface before filling. Never try to fill a
cassette while it is in the machine.
1
2
AB.60.054
Filling with Easy-Fil system
The Easy-Fil system consists of three elements:
•
The bottle collar.
•
The bottle adapter.
•
The agent filling port.
CAUTION
WARNING
13 Vaporizer cassettes
2067226-001
13-11
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...