Changing a cassette during a case
1.
Push the Agent quick key.
2.
Set agent delivery to off.
3.
Turn the lock on the handle to the vertical position (Aladin2
cassettes only).
4.
Remove cassette from the active bay.
•
Store the cassette in the cassette storage bay if it will not be
used.
•
Fill the cassette following the appropriate filling procedure if
the cassette requires filling.
5.
Install a cassette following the Installing a cassette procedure.
A tone sounds.
The agent setting in the quick key area flashes.
If cassette of same agent type is installed during a case, the last
agent setting used shows the agent quick key.
6.
Change or confirm the agent setting.
7.
To confirm the agent setting:
•
Push the ComWheel to confirm the setting.
•
Push the Agent quick key to confirm the setting.
If the setting is not confirmed within 30 seconds, the agent
delivery is set to off.
8.
To change an agent setting:
•
Turn the ComWheel to the desired setting.
•
Push the ComWheel or push the quick key to confirm the
setting.
Aisys CS²
13-6
2067226-001
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...