Ventilator accuracy data
The following accuracy data is based on patient conditions and
settings described in ASTM F1101. The ventilator is assumed to be
operating in volume mode. For the following to be true, the ventilator
is operating with 100 percent oxygen in the breathing system, or it is
connected to an anesthesia gas analyzer. If the ventilator is
operating without being connected to an anesthesia gas analyzer, or
at a temperature that differs from when it was calibrated, additional
errors may occur.
The minimum detectable breath size is 5.0 ml.
.
Delivery accuracy
Volume delivery accuracy
Greater than 210 ml tidal volume - accuracy better than 7%.
Less than or equal to 210 ml but greater than or equal to 60 ml tidal
volume - accuracy better than 15 ml.
Less than 60 ml tidal volume - accuracy better than 10 ml.
Volume monitoring accuracy
Greater than 210 ml tidal volume - accuracy better than 9%.
Less than or equal to 210 ml but greater than or equal to 60 ml tidal
volume - accuracy better than 18 ml.
Less than 60 ml tidal volume - accuracy better than 10 ml.
Inspiratory pressure delivery accuracy
Greater of +/- 10% or +/- 3 cmH2O
PEEP delivery accuracy
+/- 1.5 cmH2O
Pressure monitoring accuracy
Greater of +/- 5% or +/- 2 cmH2O
Note: Gas composition errors may be in additions to the above normalized accuracy. When adding errors, positive
errors can have the effect of nulling out negative errors.
Note: Use of anesthetic agent could affect the errors by approximately -0.95%/% volume agent in normal mode.
Note: Temperatures greater than or less than 24°C may decrease BTPS accuracy.
AC.22.029
Figure 11-11 • Gas composition related errors
11 Specifications and theory of operation
2067226-001
11-39
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...