1
2
T
I
3
4
6
5
AB.91.040
1.
Mandatory SIMV breath
2.
Spontaneous pressure supported breath
3.
Paw
4.
Psupport
5.
PEEP
6.
Time
Figure 11-6 • SIMV VCV waveform
SIMV PCV mode
Synchronized intermittent ventilation with pressure control ventilation
(SIMV PCV) delivers a relatively slow breathing rate with pressure-
controlled breathing. This mode combines mandatory breaths with
spontaneous breath support. If a trigger event occurs within the
synchronized window, a new pressure-controlled breath is initiated. If
a trigger event occurs elsewhere during the expiratory phase, a
support for a spontaneous breath is provided with pressure support
added as set by the clinician.
SIMV PCV mode settings:
•
Pinsp
•
RR
•
Tinsp
•
Psupport
•
PEEP
•
Pmax
•
Trig Window
•
Flow Trigger
•
End of Breath
•
Rise Rate
•
Exit Backup
11 Specifications and theory of operation
2067226-001
11-29
Summary of Contents for auisys cs2
Page 1: ...Aisys CS User s Reference Manual Software Revision 10 X ...
Page 32: ...1 20 2067226 001 ...
Page 60: ...3 Set the alarm limits to clinically appropriate settings Aisys CS 3 12 2067226 001 ...
Page 87: ...3 Push the drain button to empty any water in the condenser 3 Operation 2067226 001 3 39 ...
Page 90: ...3 42 2067226 001 ...
Page 94: ...4 4 2067226 001 ...
Page 132: ...7 20 2067226 001 ...
Page 164: ...9 8 2067226 001 ...
Page 178: ...System pneumatic circuit AB 75 270 Aisys CS 11 2 2067226 001 ...
Page 230: ...4 Select Back to return to the Gas Usage menu Aisys CS 12 4 2067226 001 ...
Page 262: ...13 20 2067226 001 ...
Page 268: ...Aisys CS User s Reference Manual English 2067226 001 06 13 C 18 05 02 Printed in USA ...