Visually check the cooling system for leakages.
N
No
otte
e
Do not mistake condensation for leakage.
1.
Remove the rear centre hatch to gain access to the
internal cooling system.
2. Inspect the internal cooling system for visible leaks.
Pay special attention to connectors.
3. Verify the head pressure is set to 1
1 b
ba
arr ±
±0
0..1
1 on the
manometer placed close to the expansion chamber.
If the pressure deviates by more than
±
±0
0..1
1 from 1
1 b
ba
arr, top up media according to Mainte-
nance Manual.
4. Press M
Me
en
nu
u.
5. Press S
Se
errvviicce
e P
Pa
ag
ge
e.
6. Press H
Ha
arrd
dw
wa
arre
e S
Stta
attu
uss.
7. Verify that the indicator
Head pressure
in EBM Con-
trol corresponds with the pressure indicated on the
manometer.
01
IM-710429
en-GB
77
Summary of Contents for ArcamEBM Spectra H
Page 1: ...Original language 01 IM 710429 en GB I IM M Installation Manual Spectra H...
Page 2: ...This page is intentionally left blank...
Page 4: ...Table of contents ii 01 IM 710429 en GB This page is intentionally left blank...
Page 6: ...Table of contents ii 01 IM 710429 en GB This page is intentionally left blank...
Page 8: ...This page is intentionally left blank 2 Revision Notes 4 01 IM 710429 en GB...
Page 36: ...This page is intentionally left blank 5 Safety 32 01 IM 710429 en GB...
Page 44: ...This page is intentionally left blank 6 Technical Data 40 01 IM 710429 en GB...