It is recommended that the air supply shut-off valve is located m
ma
axxiim
mu
um
m 5
5 m
me
ette
errss away from the machine.
P
Pu
urriittyy C
Clla
assss
3-5-3 (ISO 8573-1)
Demands normally reached in systems where bulk liquid is removed via an air
dryer.
M
Miin
niim
mu
um
m ssu
up
pp
pllyy
p
prre
essssu
urre
e
5 bar (73 PSI)
R
Re
ecco
om
mm
me
en
nd
de
ed
d ssu
up
p--
p
pllyy p
prre
essssu
urre
e
6 bar (90 PSI)
M
Ma
axxiim
mu
um
m ssu
up
pp
pllyy
p
prre
essssu
urre
e
9 bar (130 PSI)
M
Miin
niim
mu
um
m a
aiirr ffllo
ow
w
d
de
em
ma
an
nd
d
EBM Machine: 20 l/min at 6 bar
Powder Recovery Station: 1200 l/min at 6 bar
Vacuum Cleaner: 630 l/min at 6 bar
A higher air flow is required when the recycle vacuum cleaner is running.
N
No
otte
e
A reduced suction performance in the vacuum cleaner might occur if the flow is too low.
01
IM-710429
en-GB
37
Summary of Contents for ArcamEBM Spectra H
Page 1: ...Original language 01 IM 710429 en GB I IM M Installation Manual Spectra H...
Page 2: ...This page is intentionally left blank...
Page 4: ...Table of contents ii 01 IM 710429 en GB This page is intentionally left blank...
Page 6: ...Table of contents ii 01 IM 710429 en GB This page is intentionally left blank...
Page 8: ...This page is intentionally left blank 2 Revision Notes 4 01 IM 710429 en GB...
Page 36: ...This page is intentionally left blank 5 Safety 32 01 IM 710429 en GB...
Page 44: ...This page is intentionally left blank 6 Technical Data 40 01 IM 710429 en GB...