Diagnostics
6-5
September 2004
Deactivating the Tracking
Window
1.
Press
[ESC]
,
[ENTER]
,
[ESC]
. The meter enters the user program.
2.
Press [
X
] until
PROG
appears and press
[ENTER]
.
3.
Press [
X
] until
CH1
or
CH2
appears and press
[ENTER]
.
4.
Press [
X
] until
PIPE
appears and press
[ENTER]
.
5.
Press [
ENTER
] until
TRACKING WINDOWS
appears.
6.
Select
NO
to deactivate the Tracking Window.
Entering the Constant
Sound Speed
1.
At
FLUID TYPE
, select
OTHER
.
2.
At
FLUID SOUNSPEED
, enter the measured sound speed and press
[
ENTER
]. Press [
W
]
and [
X
] to move the cursor to desired
location and press
[
S
]
and
[
T
]
to increase or decrease the
number and press
[ENTER].
3.
Press [
ESC
] three times to exit the
User Program
.
4.
The AT868 returns to making measurements. If desired,
reconfigure the display to show the desired measurement as
discussed on the previous page.
Diagnostic Record
The values for the diagnostic parameters immediately after initial
installation of the meter and verification of proper operation should
be entered in Table E-2 in Appendix E,
. These initial
values can then be compared to current values to help diagnose any
future malfunction of the system.
Summary of Contents for AquaTrans AT868
Page 8: ...Chapter 1 ...
Page 24: ...Chapter 2 ...
Page 60: ...Chapter 3 ...
Page 66: ...Chapter 4 ...
Page 72: ...Chapter 5 ...
Page 77: ...Chapter 6 ...
Page 88: ...Chapter 7 ...
Page 98: ...Chapter 8 ...
Page 106: ...Appendix A ...
Page 112: ...Appendix B ...
Page 113: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 LVD Compliance B 2 ...
Page 116: ...Appendix C ...
Page 117: ...Data Records Site Data C 1 ...
Page 122: ...Appendix D ...
Page 123: ...Measuring P and L Dimensions Measuring P and L D 1 ...
Page 126: ...Appendix E ...
Page 127: ...Service Record Introduction E 1 Data Entry E 1 Diagnostic Parameters E 3 ...
Page 131: ...Appendix F ...