Diagnostics
6-1
September 2004
Introduction
This chapter explains how to troubleshoot the Model AT868 if
problems arise with the electronics enclosure, the flowcell, or the
transducers. Indications of a possible problem include:
•
display of an error message on the LCD screen
•
erratic flow readings
•
readings of doubtful accuracy (i.e. readings that are not consistent
with readings from another flow measuring device connected to
the same process).
If any of the above conditions occurs, proceed with the instructions
presented in this chapter.
Displaying Diagnostic
Parameters
The Model AT868 has built-in
Diagnostic Parameters
to aid in the
troubleshooting of flowcell, transducer and/or electrical problems. To
access these parameters, do the following (refer to Figure A-2 on
page A-2 in Appendix A,
):
Accessing the Display
Submenu
1.
Press
[ESC]
,
[ENTER]
,
[ESC]
. The meter enters the user program.
2.
Press [
X
] until
PROG
appears and press
[ENTER]
.
3.
Press [
X
] until
GLOBL
appears and press
[ENTER]
.
4.
Press [
X
] until
I/O
appears and press
[ENTER]
.
5.
Press [
X
] until
LCD
appears and press
[ENTER]
.
Configuring the Display
1.
At
# of LCD PARAMS
, press [
X
] to select the desired number of
parameters to be sequentially displayed and press
[ENTER]
.
For a 1-channel/path AT868, proceed to Step 3. For a 2-channel/path
meter, proceed to Step 2.
Note:
If you select
OFF
, the meter returns to the
Global I/O
window.
Go to
Procedure Options
on page 6-3.
2.
Press [
X
] to select the desired channel/path option and press
[ENTER]
.
3.
At
Measurement Name
, press [
X
] to select the desired parameter
(refer to Table 6-1 on the next page for a list of available
diagnostic parameters) and press
[ENTER]
.
Summary of Contents for AquaTrans AT868
Page 8: ...Chapter 1 ...
Page 24: ...Chapter 2 ...
Page 60: ...Chapter 3 ...
Page 66: ...Chapter 4 ...
Page 72: ...Chapter 5 ...
Page 77: ...Chapter 6 ...
Page 88: ...Chapter 7 ...
Page 98: ...Chapter 8 ...
Page 106: ...Appendix A ...
Page 112: ...Appendix B ...
Page 113: ...CE Mark Compliance Introduction B 1 EMC Compliance B 1 LVD Compliance B 2 ...
Page 116: ...Appendix C ...
Page 117: ...Data Records Site Data C 1 ...
Page 122: ...Appendix D ...
Page 123: ...Measuring P and L Dimensions Measuring P and L D 1 ...
Page 126: ...Appendix E ...
Page 127: ...Service Record Introduction E 1 Data Entry E 1 Diagnostic Parameters E 3 ...
Page 131: ...Appendix F ...