Checkout
2001989-351A
ApexPro™
7-7
2.
Press
Enter
and minimize the MS-DOS screen.
3.
At a CIC, right-click on an empty available bed. Select newly admitted test bed
and verify ECG waveform appears without dropout.
4.
Restore the MS-DOS window either at the CIC, or return to your ApexPro
Telemetry System and restore the MS-DOS window there.
5.
Type
discharge <unit>|<bed>
and press
Enter
.
6.
At the message,
Are you sure you want to delete?
, type
y
.
7.
Repeat these steps for all remaining receivers.
Electrical safety tests
Electrical safety tests provide a method of determining if potential electrical health
hazards to the patient or operator of the device exist.
Recommendations
Test conditions
Electrical safety tests may be performed under normal ambient conditions of
temperature, humidity, and pressure.
Test equipment
The recommended test equipment required to perform electrical safety tests is listed
below.
Power outlet test
Before starting the tests, the power outlet from which the monitoring device will get
electrical power must be checked. This test checks the condition of the power outlet
to ensure correct results from leakage tests.
For international power outlets, refer to the internal standards agencies of that
particular country. Use a digital multimeter to ensure the power outlet is wired
properly.
If other than normal polarity and ground is indicated, corrective action must be taken
before proceeding. The results of the following tests will be meaningless unless a
properly wired power outlet is used.
Item
Specification
Leakage Current Tester
Equivalent to the circuits shown
Digital Multimeter (DMM)
AC volts, ohms
Ground Bond Tester
0 – 1 ohm
ECG Test Body
All leads together
Summary of Contents for ApexPro TT
Page 9: ...2001989 351A ApexPro vii...
Page 10: ...viii ApexPro 2001989 351A...
Page 11: ...2001989 351A ApexPro 1 1 1 Introduction...
Page 20: ...1 10 ApexPro 2001989 351A Introduction...
Page 21: ...2001989 351A ApexPro 2 1 2 Equipment Overview...
Page 41: ...2001989 351A ApexPro 3 1 3 Installation and configuration...
Page 62: ...3 22 ApexPro 2001989 351A Installation and configuration...
Page 63: ...2001989 351A ApexPro 4 1 4 Maintenance...
Page 84: ...4 22 ApexPro 2001989 351A Maintenance...
Page 85: ...2001989 351A ApexPro 5 1 5 Troubleshooting...
Page 109: ...2001989 351A ApexPro 6 1 6 Replaceable parts...
Page 138: ...6 30 ApexPro 2001989 351A Replaceable parts Quad receiver module...
Page 139: ...2001989 351A ApexPro 7 1 7 Checkout...
Page 163: ...2001989 351A ApexPro A 1 A Technical specifications...
Page 179: ...2001989 351A ApexPro B 1 B Electromagnetic compatibility...
Page 190: ...B 12 ApexPro 2001989 351A Electromagnetic compatibility...
Page 191: ...2001989 351A ApexPro 1 1...