5 Setups, alignments and adjustments
32551-IMG rev 6
GE Healthcare
47
5 Setups, alignments and adjustments
5.1
S
ETUP
,
ALIGNMENT
AND
ADJUSTMENT
PRIOR
TO
USE
5.1.1
Initial setups
In this section the contents of the nonvolatile memory are checked and
changed if necessary. The setup information is available in two programs,
PAr
and
SUP
. The automatic exposure control information is available in
AEC
and
tAr
programs.
In the SetUP (prog. #52 SUP) the following is setup:
In the PArameter (prog. #53 PAr) the following is setup:
In the Automatic Exposure Control (prog. #64 AEC) information is
available:
In the tArget time (prog. #65 tAr) following information is available:
#52 SUP
A
Compression autorelease time
#52 SUP
L
Label type
#52 SUP
H
Label text
#52 SUP
i
i/o device
#52 SUP
F
Default filter
#53 PAr
S
The serial number of the unit
#53 PAr
C
Country code
#53 PAr
r
Tube efficiency in mR/mAs for both filters
#53 PAr
b
Anode brake selection
#53 PAr
t
Software mode and Phototimer modes
#53 PAr
I
Collimator type (Interlock)
#53 PAr
d
C-arm control board / CD board selection
#64 AEC
d
Wanted optical density
#64 AEC
h
IQC - kV programming
#64 AEC
I
Target time MIN/MAX automatic exposure
#64 AEC
c
Thickness compensation adjustment
#64 AEC
g
AEC gain - optical density programming
#64 AEC
E
Full AEC kV programming (Initial kV vs. breast
thickness)
#64 AEC
U
kV compensation adjustment
#64 AEC
S
Density step adjustment
#65 tAr
The Alpha RT AA- mode (auto kV) selects the used kV
according to a
target time
which is programmed for small
and large
focus
, and for Mo and Rh
filters
separately. The
target time depends on the compressed breast thickness
and is a linear curve calculated from two programmed
points.
Summary of Contents for Alpha RT 32551
Page 1: ...GE Healthcare Alpha RT Installation Instructions 32551 IMG rev 6 0459 ...
Page 2: ......
Page 4: ......
Page 42: ...3 Installation 36 GE Healthcare 32551 IMG rev 6 ...
Page 51: ...4 Alpha RT softstart procedure 32551 IMG rev 6 GE Healthcare 45 Fig 4 10 Inverter board ...
Page 52: ...4 Alpha RT softstart procedure 46 GE Healthcare 32551 IMG rev 6 ...
Page 71: ...5 Setups alignments and adjustments 32551 IMG rev 6 GE Healthcare 65 Fig 5 2 Magnetic brake ...
Page 94: ...5 Setups alignments and adjustments 88 GE Healthcare 32551 IMG rev 6 ...
Page 96: ...6 Alpha RT installation step by step 90 GE Healthcare 32551 IMG rev 6 ...
Page 97: ......
Page 98: ......