9.9.2
Manual instructions - Flow path
The table below describes the flow path related manual instructions available for ÄKTA pure.
Description
Instruction name
Inlet A turns Inlet valve A to the selected position.
Note:
Positions A1 - A7 are available, if Inlet valve V9-IA or V9H-IA (7-
ports) is used. Only positions A1 and A2 are available, if Inlet valve
V9-IAB or V9H-IAB (2-ports) is used.
Inlet A
Inlet B turns the Inlet valve B to the selected position.
Note:
Positions B1 - B7 are available, if Inlet valve V9-IB or V9H-IB (7-
ports) is used. Only positions B1 and B2 are available, if Inlet valve
V9-IAB or V9H-IAB (2-ports) is used.
Inlet B
Sample inlet turns the Sample inlet valve to the selected position.
Sample inlet
Injection valve sets the Injection valve to the selected position.
The instruction gives an injection mark in the chromatogram when
the inlet valve switches to Inject or Direct Inject.
Note:
Sample flow refers to the flow that enters the injection valve via
SaP port.
Injection valve
Column position turns the Column valve to the position specified
in the parameter Position.
Column position
pH valve sets the pH cell and the 0.2 MPa restrictor in positions
inline or offline.
pH valve
The pH valve also has a calibration position. This position is only
available when performing calibration of the pH monitor (In Sys-
tem control select System:Calibrate).The calibration position can
also be used to fill the pH cell with storage solution since the pH
valve is in open position.
Note:
• It is not possible to turn the pH valve during any type of fraction-
ation as it affects the delay volume.
• The pH valve instruction can be given during the delay volume
of the different stop fractionation instructions, but it is executed
only after the set delay volume has been collected.
Outlet valve turns the Outlet valve to the selected position. The
instruction gives a mark in the chromatogram when the valve is
switched to the selected position.
Outlet valve
ÄKTA pure User Manual 29119969 AB
471
9 Reference information
9.9 Manual instructions
9.9.2 Manual instructions - Flow path
Summary of Contents for AKTApure
Page 1: ...ÄKTA pure User Manual ...
Page 2: ...Page intentionally left blank ...
Page 520: ...Page intentionally left blank ...
Page 521: ...Page intentionally left blank ...