Meters and sensors
The following table lists the meters and sensors that are shown in the flow chart.
Note
Function
Tag
Movable
Buffer inlet air sensor
AE-151
Final check that no air enters the column
Pre-column air sensor
AE-152
Post-column pH-meter
AT-121
Peak detection
Post-column UV-meter
AT-131
Also includes a temperature meter
Pre-column conductivity meter
CE/TE-101
Peak detection and CIP-control, also includes
a temperature meter
Post-column conductivity
meter
CE/TE-102
Measures the total system flow
System flow meter
FT-141
Air trap high level meter
LEH-167
Air trap low level meter
LEL-166
Option
Pre-filter pressure meter
PT-111
Guards the column from over pressure,
detects clogged column
Pre-column pressure meter
PT-112
Part of sample pump option
Sample pump pressure meter
PT-114
part of PCV option
PCV pressure meter, A inlets
PT-115
part of PCV option
PCV pressure meter, B inlets
PT-116
76
ÄKTAprocess Operating Instructions 29-0252-49 AA
3 System description
3.5 Flowchart