Temperature and pressure limits
The table below gives temperature and pressure limits for various types of tubing.
Maximum pres-
sure
Fluid temperature
Tubing diameter
Tubing material
6 bar*
4ºC to 40ºC
All
Polypropylene
3 bar
40ºC to 60ºC
10 bar*
4ºC to 40ºC
3/8", 1/2"
Stainless steel
6 bar*
1"
3 bar
40ºC to 60ºC
All
1 bar
60ºC to 80ºC
Footnote
* When a capsule filter is used the maximum pressure is 5 bar.
System volume
The values presented in this table represents approximate hold-up volumes for indicated
sections of the system. Values may vary between individual systems.
System tubing diameter
Flow path section
1"
10 mm to 1/2"
6 mm to 3/8"
2250 ml
550 ml
300 ml
From the inlets to the column
(Air trap, Filter and Column(s)
bypassed)
590 ml
47 ml
22 ml
From the column to the UV cell
350 ml
33 ml
23 ml
From the UV cell to the outlets
1800 ml
1200 ml
1200 ml
Air trap
2700 ml
450 ml
440 ml
Filter housing (empty)
166
ÄKTAprocess Operating Instructions 29-0252-49 AA
8 Reference information
8.1 Specifications