3.4
Equipment
Pinch valves
Pinch valves are used to control the path of the
liquid flow through the system. There are 19 pinch
valves in total. The valves have manually operated
safety locks keeping the tubing in place and pre-
venting the valves from accidentally closing during
flow kit installation.
WARNING
Never put fingers or any objects other than the intended tubing in
to the pinch valve opening.
Default position
Valves
Closed
Inlet valves, XV-001 to XV-006
Inline
Air trap valves, XV-021 to XV-023
Bypass
Column valves, XV-031 to XV-033
Closed
Outlet valves, XV-051 to XV-056
Closed
Air Vent Valve, HV-301
During pause, the air trap valves will remain in the same position as during the run. The
user can change this by setting the air trap valves to inline during pause, in the System
Settings in UNICORN.
48
ÄKTA ready Operating Instructions 28960345 AD
3 System description
3.4 Equipment