Conditional fractionation
The Conditional Fractionation phase sets conditions for fractionation start and stop
that will apply in subsequent phase(s). The conditions will apply until fractionation is
stopped, either because the Stop Condition is met or the Stop fractionation at the end
of this phase is checked in a subsequent phase.
Conditional Fractionation is not included in predefined methods but can be
added if required from the phase library.
Note:
Conditional Fractionation can be used to set conditions for fractionation start and stop
based on one signal (e.g., conductivity) while performing peak fractionation based on
another signal (e.g., UV absorbance).
To use Conditional Fractionation, insert the phase before the one where the conditions
are required. Set the fractionation settings in the Conditional Fractionation phase, and
set the next phase to Continue ongoing.
To stop conditional fractionation, include a Fractionation Stop phase from
the phase library.
Note:
ÄKTA pilot 600 User Manual 29274559 AA
75
4 Operation
4.2 Fractionation
Summary of Contents for AKTA pilot 600