The four options for unpacking are used to simplify unpacking of columns. Before using
an unpacking option, make sure to read each column instruction.
Column pressure monitoring
The pre- and post-column pressure sensors are located at the inlet and outlet of the
column valve.
The pressure drop between a sensor and a column is automatically compensated. This
affects pressures at higher flow rates (>600 mL/min).
There are four pressure signals, see the image and table below.
A
B
C
D
Purpose
Raw/compensated
Signal
Part
Used to protect the column
hardware
Compensated signal
Pre-column pressure
A
Used to protect the column
bed
Compensated signal
Delta column pressure
B
For reference
Raw signal
Pre-column sensor
pressure
C
Used to protect the pH
electrode and the UV cell
Raw signal
Post-column sensor
pressure
D