Action
Description
Error
code
Restart the instrument. If recurrent please
contact Service.
Lamp error
28, 29
See UV monitor U9-M and UV detector
unit, on page 363.
High temperature
30
UV Detector U9-D
Action
Description
Error
code
Restart the instrument. If recurrent please
contact Service.
Internal instrument error
24, 26, 28,
29, 31, 33
Check that the detector and flow cell are
fitted correctly. If warning reappears,
contact Service.
Too low light intensity
25
Check that the detector and flow cell are
fitted correctly. If warning reappears,
contact Service.
Too high light intensity
27
Check that the detector is fitted correctly.
If warning reappears, contact Service.
Too high light intensity, R
channel
30
Check that the flow cell is fitted correctly.
If warning reappears, contact Service.
Too high S light intensity, S
channel
32
Check optical pathway and restart the
instrument. If recurrent please contact
Service.
No light detected
34
Check that the detector is fitted correctly.
If warning reappears, contact Service.
Too low light intensity, R
channel
35
No light through flow cell. Check solution
absorption and that the cell is fitted cor-
rectly.
Too low light intensity, S
channel
36
418
ÄKTA avant User Manual 29035184 AE
8 Troubleshooting
8.9 Error codes