Workflow for preparing the
built-in fraction collector
Before starting to prepare the built-in fraction collector, check the fractionation settings
in the method to be run. Perform the steps described below according to the settings in
the method.
Reference to instructions
Operator actions
Step
See Prepare and insert the cassette
tray below.
Insert the cassette tray or a rack for
tubes or bottles.
1
See UNICORN System Control
Manual.
Change the System Settings in
UNICORN to set the fractionation
mode and other settings for fraction
collection.
2
Available System Settings are de-
scribed in Section 9.7.7 System
settings - Fraction collection, on
page 468.
Prepare cassettes and insert the
cassette tray
Follow the instructions to prepare the fraction collector before a run.
Action
Step
If you are to use cassettes with the QuickRelease function, first open the
cassettes.
1
204
ÄKTA avant User Manual 29035184 AE
5 Operation
5.8 Fractionation
5.8.1 Prepare the built-in fraction collector
Summary of Contents for AKTA avant
Page 1: ...ÄKTA avant User Manual ...
Page 2: ...Page intentionally left blank ...
Page 536: ...Page intentionally left blank ...
Page 537: ...Page intentionally left blank ...