•
Spread of biological agents. The operator has to take all necessary actions to avoid
spreading hazardous biological agents in the vicinity of the instrument. The facility
should comply with the national code of practice for biosafety.
•
High pressure. The ÄKTAavant instrument operates under high pressure. Wear
protective glasses at all times.
Cautions
CAUTION
To avoid hazardous situations when working with the ÄKTAavant
system, take the following measures for personal protection.
•
Close doors. To minimize the risk of exposure to hazardous chemicals and pressurized
liquids, always close the Foldable door and the Pump cover before starting a run.
•
Cut injuries. The tubing cutter is very sharp and must be handled with care to avoid
injuries.
Installing and moving the
instrument
Warnings
WARNING
To avoid personal injury when installing or moving the ÄKTAavant
instrument, follow the instructions below.
•
Heavy object. The ÄKTAavant instrument weighs about 100 kg. Use proper lifting
equipment, or use four or more persons when moving the instrument. All lifting and
moving must be performed in accordance with local regulations.
•
Moving the instrument horizontally. Three persons are recommended when moving
the instrument horizontally.
•
Supply voltage. Make sure that the supply voltage at the wall outlet corresponds to
the marking on the instrument, before connecting the power cord.
•
Protective ground. The ÄKTAavant instrument must always be connected to a
grounded power outlet.
•
Power cord. Only use power cords delivered or approved by GE Healthcare.
Getting Started with ÄKTAavant and UNICORN 6 28-9440-69 AB
15
2 Safety instructions
2.1 Safety precautions
Summary of Contents for AKTA avant
Page 1: ...GE Healthcare Getting Started with KTAavant and UNICORN 6...
Page 2: ......
Page 133: ......