General Procedures and Functional Checks
4-10
LOGIQ C9 Series – Basic Service Manual
5482298-100 English Rev. 5
LCD Monitor position adjustment
Refer to User Manual for LCD Monitor position adjustment.
Logging on to LOGIQ C9 Series as “ADM”
Select
Utility
tab on the touch screen, then select
Config
.
It will bring up the
Operator Login
dialog where you must log
on.
Figure 4-3. Operator Login Window
1.
Operator
: Select the operator.
2.
Password
: Enter Operator’s password (optional).
3. Select the type of Login or Cancel.
•
OK
: Standard login.
•
Cancel
: Cancel login.
As default, two users are defined, USR and ADM.
• USR
If you log on as
USR,
you will have access to do set-up tasks
that a user may need to do during daily use.
As default, no password has been set for
USR
. Just type the
name
USR
, and select
Login
.
• ADM
If you log on as
ADM
, you will have access to do general set-up,
service adjustments, adjust network and connectivity settings.
As default, the password for
ADM
is
ulsadm
.
Select the name
ADM
, the password (
ulsadm
) and select
Login
.
It is possible for the administrator (
ADM
) to establish new users
and set unique passwords for each user, including a new
password for ADM. If the login as ADM fails, contact the
responsible person in the hospital to get access.
Summary of Contents for 5489329
Page 5: ...LOGIQ C9 Series Basic Service Manual i 3 5482298 100 English Rev 5 ...
Page 6: ...i 4 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 7: ...LOGIQ C9 Series Basic Service Manual i 5 5482298 100 English Rev 5 ...
Page 8: ...i 6 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 9: ...LOGIQ C9 Series Basic Service Manual i 7 5482298 100 English Rev 5 ...
Page 10: ...i 8 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 11: ...LOGIQ C9 Series Basic Service Manual i 9 5482298 100 English Rev 5 ...
Page 12: ...i 10 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 13: ...LOGIQ C9 Series Basic Service Manual i 11 5482298 100 English Rev 5 ...
Page 18: ...i 16 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 52: ...Introduction 1 28 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 76: ...Site Preparations 2 24 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 118: ...System Setup 3 42 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 206: ...Renewal Parts 9 8 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 242: ...Care and Maintenance 10 36 LOGIQ C9 Series Basic Service Manual 5482298 100 English Rev 5 ...
Page 247: ...GE ...