Configuration:
Service Mode settings
2000966-542D
Dash 3000/4000/5000
4-13
Admit Menu
The Admit Menu selection determines the function of the patient
monitor. Before configuring the Admit Menu, you must know if the
patient monitor will be used for standard adult, neonatal, or operating
room monitoring, and if the patient monitor will be moved from room to
room. All combinations are explained below.
STANDARD
—configures the patient monitor to stay in one room for
stationary monitoring only. Patient monitors not connected to the
CARESCAPE Network (Ethernet connection) must use
STANDARD
configuration only.
ROVER
—configures the patient monitor to move from room to room
for stationary monitoring only.
COMBO
—configures the patient monitor to stay in one room for
both stationary and ambulatory (telemetry) monitoring. This patient
monitor displays all Tram module data combined with ECG data for
ambulatory patients.
NOTE
Users should be aware of a possible time discrepancy between
the waveforms from the Telemetry device and the waveforms
hard-wired to the Dash patient monitor. Users should not
consider these waveforms to be synchronous. If absolute
synchronicity is desired, Combo mode should be discontinued and
the ECG waveforms should be acquired via the hard-wired
bedside patient monitor.
ROVER COMBO
—configures the patient monitor to move from
room to room for both stationary and ambulatory (telemetry)
monitoring.
1. Enter service mode (See
“Service Mode settings” on page 4-11
) and
select
MENU SETUP
>
ADMIT MENU
.
2. Use the
Trim Knob
control to select the function of the patient
monitor.
3. Press
Trim Knob
control to exit.
↓
↑
513A
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......