Installation:
Connections
2000966-542D
Dash 3000/4000/5000
3-7
Defib sync
The connector provides ECG analog output signals to user-supplied
equipment.
CAUTION
Equipment damage. Connect all peripheral equipment
before plugging the power cord into an AC outlet.
Otherwise, connectors may be damaged.
AC power
Use this connector to apply power to the patient monitor. The patient
monitor will be powered at all times when using AC power (there is no
AC power switch). The patient monitor is preset at the factory for a
specific line voltage. Before applying power, verify the patient monitor’s
line voltage setting matches your power supply. If required, use a small
bladed screw driver to change the line voltage setting. Refer to the label
on the back of the unit for the voltage and current requirements.
CAUTION
Verify the line voltage selector setting before plugging
the power cord into an AC outlet. The power supply may
be damaged if the power source applied is different from
the line voltage setting.
Power up
After making all connections, plug the power cord into an AC wall outlet.
When all cables are properly connected, press the
Power
button to turn
the patient monitor on. All front panel indicators will illuminate until
the power-up sequence is complete. After approximately 20 seconds you
should see a display on the screen.
Configure
Go to
on page 4-1 to configure and set up the Dash
patient monitor for patient use.
115V
line voltage selector:
115V for a 100-120V power source
230V for a 220-240V power source
004B, 829A
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......