Equipment overview:
Components
2-10
Dash 3000/4000/5000
2000966-542D
Verify compatibility of an unmarked battery as follows.
1. Install a battery pack in the patient monitor.
2. Using the
Trim Knob
control, access the Service Mode
menu
starting from the Main
Menu. Select
MORE MENUS
>
MONITOR
SETUP
>
SERVICE MODE
.
3. Enter password using the
Trim Knob
control to select the day and
month from patient monitor screen with leading zeros. (e.g. July 4 =
0407).
4. Select
BATTERY SERVICE
.
5. Verify that the
MANUFACTURER NAME
does not display
INCOMPAT
,
NME
, or
UNKNOWN
for the battery corresponding to
BATTERY A
or
BATTERY B
slot.
Optional components
TRAM-rac 2A module housing
The TRAM-rac 2A module housing currently supports the SAM and ICG
modules.
An integral power supply is used to run the TRAM-rac 2A and support
the needed voltages.
Dash Port 2 docking station
The docking station is a quick mount/dismount base for a Dash patient
monitor. It gives the patient monitor easy connect or disconnect access to
AC power, CARESCAPE Network, a remote display, and auxiliary
devices.
See the Dash Port 2 Docking Station Operating Instructions and the
Dash Port 2 Docking Station Service Manual for additional information.
NOTE
When a Dash patient monitor is connected to the docking station,
only the docking station’s Ethernet port is active. The Dash patient
monitor’s network port remains inactive until the patient monitor is
disconnected from the docking station.
797B
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......