Functional and electrical safety checks:
Functional Checkout procedures
8-32
Dash 3000/4000/5000
2000966-542D
Pulse oximetry tests for Masimo SET SPO2
Equipment
Use the following equipment for these tests:
Masimo SpO
2
Test Kit (2021087-001). Includes Masimo Tester and
SPO2 Sensor Adapter Cable.
NOTE
The Masimo Tester is intended for use only with Masimo
instruments or instruments containing Masimo SET oximetry
with standard Masimo PC Series patient cable connectors. See
the Masimo “Directions for Use” that came with the Tester for
more information
The PPR and SPO2 values for the Masimo Tester must be within
these limits: Peripheral pulse rate: 61 bpm
1 bpm; and SPO2
value: 81%
3%.
Procedures
1. Connect the Masimo-style SPO
2
simulator cable between the SPO2
connector of the patient monitor and the SPO
2
simulator.
2. Turn on the patient monitor.
3. Verify the following are displayed at the patient monitor: (It might be
necessary to turn the SPO
2
parameter on.)
A waveform with an SpO
2
label.
An SPO
2
% reading between
78 - 84
%.
A PPR reading between
60
and
62
beats per minute.
4. Disconnect the simulator cable from the module.
824C
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......