Field replaceable units:
Replace display flex assembly
7-24
Dash 3000/4000/5000
2000966-542D
Replace display flex assembly
Dash 3000 display flex assembly
The Dash 3000 display assembly includes the flex assembly.
1. Remove the handle and display assemblies according to steps
starting on page
.
2. Replace the display assembly and reassemble the patient monitor in
reverse order of disassembly.
Dash 4000 and 5000 display flex assembly
CAUTION
FLEX CABLE DAMAGE —
Do not
kink, pinch, stretch,
twist, or tightly fold a flex cable.
1. Loosen the handle assembly as required. Refer to
NOTE
This step is not necessary for Dash 5000 patient monitors.
2. Remove the display assembly from the main unit. Refer to page
NOTE
Use a thick book or some heavy object to support the display
assembly while you attach the display flex to the main unit.
3. Remove the two screws to disconnect the flex cable from the display.
Retain the screws.
4. Insert the new flex cable connector into the display connector as
shown in the figure below.
5. Using two screws, secure the flex cable to the display assembly.
873A
Summary of Contents for 001C 051D 003A
Page 10: ...viii Dash 3000 4000 5000 2000966 542D ...
Page 11: ...2000966 542D Dash 3000 4000 5000 1 1 1 Introduction ...
Page 20: ...Introduction Service information 1 10 Dash 3000 4000 5000 2000966 542D ...
Page 21: ...2000966 542D Dash 3000 4000 5000 2 1 2 Equipment overview ...
Page 64: ...Equipment overview Theory of operation 2 44 Dash 3000 4000 5000 2000966 542D ...
Page 65: ...2000966 542D Dash 3000 4000 5000 3 1 3 Installation ...
Page 73: ...2000966 542D Dash 3000 4000 5000 4 1 4 Configuration ...
Page 99: ...2000966 542D Dash 3000 4000 5000 5 1 5 Preventive maintenance ...
Page 117: ...2000966 542D Dash 3000 4000 5000 6 1 6 Troubleshooting ...
Page 146: ...Troubleshooting Wireless LAN 6 30 Dash 3000 4000 5000 2000966 542D ...
Page 147: ...2000966 542D Dash 3000 4000 5000 7 1 7 Field replaceable units ...
Page 201: ...2000966 542D Dash 3000 4000 5000 8 1 8 Functional and electrical safety checks ...
Page 257: ...2000966 542D Dash 3000 4000 5000 A 1 A Electromagnetic compatibility EMC ...
Page 267: ...2000966 542D Dash 3000 4000 5000 B 1 B Network troubleshooting ...
Page 279: ...2000966 542D Dash 3000 4000 5000 D 1 D Checklist ...
Page 281: ......