TROUBLESHOOTING: Power Source Tests
4-6
DASH 2000 Patient Monitor
Revision A
2000 412-001
Power Source Tests
AC Line Voltage Test
This test verifies that the domestic wall outlet supplying power to the
equipment is properly wired For international wiring tests, refer to the
internal standards agencies of that particular country.
120 VAC, 50/60 Hz
Use a digital voltmeter to check the voltages of the 120-volt AC wall
outlet (dedicated circuit recommended). If the measurements are
significantly out of range, have a qualified electrician repair the outlet.
The voltage measurements should be as follows:
1.
120 VAC (±10 VAC) between the line contact and neutral and
between the line contact and ground.
2.
Less than 3 VAC between neutral and ground.
240 VAC, 50/60 Hz
Use a digital voltmeter, set to measure at least 300 VAC, to check the
voltages of the NEMA 6-20R, AC wall outlet (dedicated circuit
recommended). If the measurements are significantly out of range, have
a qualified electrician repair the outlet. The voltage measurements
should be as follows:
1.
120 VAC (±10 VAC) between either ÒhotÓ contact and ground.
2.
210 to 230 VAC between the two ÒhotÓ contacts.
Neutral
Line
Ground
¨
;
¨
Hot
Hot
Ground
;
¨
¨
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......