TROUBLESHOOTING: Battery Failure
4-4
DASH 2000 Patient Monitor
Revision A
2000 412-001
Battery Failure
Defective Battery/
Battery System
The BATTERY DEFECTIVE message is displayed in the STATUS
MESSAGE line when errors have occurred within the battery
management system or the battery. The reason of the battery system
error can be found in the error log-book (MONITOR SETUP Ð>
SERVICE MENU Ð> REVIEW ERRORS Ð> VIEW OUTPUT ERRORS).
Check for PROCESS NAME: power_battm in the log-book.
The following error entries could be produced by the battery system:
Blinking Charging
Status
When the BATTERY DEFECTIVE error 0x719 occurs then additionally
the CHARGING STATUS LED is blinking yellow.
Battery Replacement
A replacement of the battery is absolutely necessary when the battery
reaches 40% full rated capacity of the 100% design capacity of the
battery pack. Then the system message REPLACE BATTERY is
displayed in the STATUS MESSAGE line. The Battery Service
Information Window in the Battery Service Menu shows the actual
FULL/NEW ratio in percent. If the usage of the Dash 2000 requires a
replacement because the FULL/NEW ratio of 40% is not acceptable then
it is up to the user to replace the battery earlier.
Follow this procedure to replace a defective battery pack in the monitor.
Preparations to Open the
Device
Before any service interventions, turn off the device and disconnect the
device from the power line. Take ESD protection precautions as
described in "Safety Information for Disassembly". Put the device on a
clean, level surface (ESD pad) which is placed on a soft material to avoid
scratches on the front panel.
0x717
No communication between PIC and PPC (12C-Error)
0x718
PIC loses communication to benchmark chip
0x719
Battery defective or disconnected or battery fuse blown
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......