Maintenance: Checkout Procedures
3-6
DASH 2000 Patient Monitor
Revision A
2000 412-001
Checkout Procedures
The following pages contain the checkout procedures for the monitor. The
purpose of the checkout procedures is to provide service personnel with a
method which can be used to verify operational and functional
performance of the monitor. Failure to attain any of the listed results
indicates a potential malfunction of the monitor.
Perform the checkout procedures upon receipt of the monitor, every twelve
months thereafter, and each time a circuit board is removed or replaced.
The checkout procedures are based on the assumption that the monitor
being tested is used with known good cables and test equipment. It also
requires that the user be somewhat familiar with the operation of all test
equipment required for the checkout procedures. For more information
concerning the operation of these components, refer to the respective
operator manual.
Manufacturer
Recommended Test
Equipment
The following table lists the manufacturerÕs recommended test
equipment, adaptors, and cables necessary to successfully complete the
checkout procedures. The checkout procedures were written for the test
equipment in the following table. If test equipment other than the
manufacturerÕs recommendation is used, it may be necessary to slightly
modify some test steps.
Monitor Power and
Battery Tests
1.
Connect a power cord between a properly wired wall receptacle and
the monitor power connector.
2.
If the unit is not turned off; press the power button to switch it off.
3.
Verify that the AC power indicator stays illuminated.
Verify that the CHARGING STATUS indicator stays illuminated
according the following list:
yellow
battery is being charged but not full
green
battery is fully charged
Description
Part Number
Qty
Multifunction Micro-simulator
MARQ1
1
Patient cable, 5-leadwire, AHA or
412931-001
1
Patient cable, 5-leadwire, IEC
412931-002
1
Leadwireset, 5-leadwire, AHA or
414556-001
1
Leadwireset, 5-leadwire, IEC
414556-003
1
BP to Simulator cable
700095-001
1
Temp to Simulator cable
6770031
1
NIBP tubing and fittings according
drawing
Digital Manometer
Sensym PDM2OOM
1
SpO
2
Simulator
408610-001
1
SpO
2
Simulator cable, Nellcor
700232-004
1
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......