
Maintenance: Maintenance Schedule
3-2
DASH 2000 Patient Monitor
Revision A
2000 412-001
Maintenance Schedule
Manufacturer
recommendation
To make sure the monitor remains in proper operational and functional
order, a good maintenance schedule must be adhered to. The
manufacturer's recommendations in this regard are as follows:
¥
Inspection
: Operators should perform this prior to admitting each
patient to the monitor. Service personnel should perform this prior to
servicing the monitor.
¥
General Cleaning
: Operators should perform this prior to
admitting each patient to the monitor. Service personnel should
perform this after servicing the monitor.
¥
Checkout Procedures
: These should be performed by qualified
service personnel upon receipt of the equipment, every 12 months
thereafter, and each time the monitor is serviced.
¥
Leakage Current Tests
: These should be performed by qualified
service personnel upon receipt of the equipment, every 12 months
thereafter, and each time the monitor is serviced.
¥
Hi-Pot Tests
: High-potential tests should be performed by qualified
service personnel whenever any component of the isolated data
acquisition system (DAS) is removed, repaired or replaced in the
monitor.
NOTE
The Hi-Pot Tests provide a means of checking the patient
isolation circuitry such that a patient receiving
defibrillation, while attached or admitted to the monitor,
will receive the full energy of each shock and that the
monitor will not absorb the energy, when delivered.
¥
Non-invasive Blood Pressure (NBP) Calibration
: NBP
calibration should be performed by qualified service personnel upon
receipt of the equipment and once each year, thereafter. Refer to
Section 5: Calibration, for this information.
Manufacturer
responsibility
Failure on the part of all responsible individuals, hospitals or
institutions, employing the use of this monitor, to implement the
recommended maintenance schedule may cause equipment failure and
potential operator and patient health hazards. The manufacturer does
not in any manner, unless an Equipment Maintenance Agreement exists,
assume the responsibility for performing the recommended maintenance
schedule. The sole responsibility rests with all individuals, hospitals, or
institutions utilizing the monitor.
Summary of Contents for marquette Dash 2000
Page 12: ...Contents viii DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 82: ...Maintenance Repair Log 3 36 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 156: ...UPPER LEVEL ASSEMBLY Spare Parts List 7 10 DASH 2000 Patient Monitor Revision A 2000 412 001 ...
Page 157: ...Revision A DASH 2000 Patient Monitor 8 1 2000 412 001 8 ASSEMBLY DRAWINGS Contents ...
Page 159: ......
Page 173: ......
Page 178: ......
Page 180: ......
Page 181: ......
Page 182: ......
Page 184: ......
Page 196: ......
Page 197: ......
Page 198: ......
Page 200: ......
Page 201: ......